State Scientific Center of the Russian Federation

20.01.2026

Вторник

14 : 01 : 30, 123

State Scientific Center of the Russian Federation

ГЭТ 17-2018 (НИЛ 2302)

Standard name:

  • ГЭТ 17-2018 State Primary Standard of Units of Dynamic and Kinematic Viscosity of Liquids

Type of measurement:

  • Measurements of the physical and chemical composition and properties of substances

Structural division: name НИО\НИЛ:

  • Laboratory of State Standards in the Field of Measurements of Liquid Density and Viscosity

Guardian Scientist:

  • Demyanov Alexey Alekseevich

Year of creation:

  • 1971

Year of re-confirmation:

  • 2018

Operating principle:

  • Viscosity measurement methods: capillary, vibration, and Stokes (rolling ball).

Composition of the standard:

  • The reference complex (EK GET 17/1-KVI) is designed to reproduce, store, and transmit the unit of kinematic viscosity in the temperature range from +20°C to +40°C;
  • The reference complex (EK GET 17/2-KVN) is designed to reproduce, store, and transmit the unit of kinematic viscosity in the temperature ranges from -40°C to +20°C and from +40°C to +150°C;
  • The reference complex (EK GET 17/3-DVP) is designed to reproduce, store, and transmit the unit of dynamic viscosity of a liquid in a flow in the temperature range from +20°C to +40°C and in the pressure range from 0.5 MPa to 4.0 MPa;
  • The reference complex (EK GET 17/4-DVD) is designed to reproduce the unit of dynamic viscosity of a liquid in the temperature range from +20°C to +40°C and in the pressure range from 0.1 MPa to 4.0 MPa.

Main metrological characteristics

Metrological characteristic name Value
Reference complex ЭК ГЭТ 17/1-КВИ
The range of values ​​of the unit of kinematic viscosity of a liquid in which the unit is reproduced, mm2/s from 4.0·10-1 to 1.0·105
Temperature range, °C from 20 to 50
The relative expanded uncertainty at the coverage factor k=2, UR, does not exceed 1,7·10-3
Reference complex ЭК ГЭТ 17/2-КВН
The range of values ​​of the unit of kinematic viscosity of a liquid in which the unit is reproduced, mm2/s from 4.0·10-1 to 1.0·105
Temperature range, °C from -40 to +150
The relative expanded uncertainty at the coverage factor k=2, UR, does not exceed 2,0·10-3
Reference complex ЭК ГЭТ 17/3-ДВП
The range of values ​​of the unit of dynamic viscosity of a liquid in a flow in which the unit is reproduced, mPa s 0,5 – 100,0
Temperature range, °C from 20 to 40
Pressure range, MPa from 0.5 to 4.0
Expanded uncertainty with coverage factor k=2, U, does not exceed, mPa s 0.25
Reference complex EK GET 17/4-DVD
Range of values ​​of the unit of dynamic viscosity of liquids in which the unit is reproduced, mPa s from 1.0 to 300.0
Temperature range, °C 20 to 40
Pressure range, MPa 0.1 to 4.0
Relative expanded uncertainty with a coverage factor of k=2 (P=0.95), UR, does not exceed 5.2 x 10-3

Verification diagram:

  • Heads the state verification scheme for liquid viscosity measuring instruments

Scope of application:

  • Oil, petrochemical, fuel and energy;
  • Aircraft engineering;
  • Food; medical, and other industries;

Results of applying the standard:

Not specified

The total number of measuring instruments traceable to ГЭТ 17-2018:

More than 100,000

Ensuring the uniformity of measurements, participation in the CIPM MRA program:

11